Home > Optical Instruments >Monochromatic Illuminator
Xenon Light Source Application Monochromatic Illuminator
 |
|
Enabling to precisely control wavelengths of output light.
- Using xenon light source 300W
- Bright monochromator (F/2.8)
- Controlling wavelength by every 1nm from 220nm to 900nm.
- Monochromatic light
(Resolution:2.6nm -) *Depends on the grating and light source
PDF brochure
|
|
System configuration example |
Component
- Xenon Light Source MAX-303
(UV, VIS and UV-VIS Mirror Module)
- Adapter lens
- Adapter cover
- Monochromator CMS-100
- System Base
- Ligth Guide
- Collimator Lens (x0.5,x1.0,x2.0)
|
Output Intensity 1.5mW/cm2 (365nm)
Measuring Condition
- MAX-303 (Including UV Mirror Module)
- CMS-100 (Blaze Wavelength:400nm)
- Quartz Light Guide (1000mm)
- Collimator Lens (x 0.5 RLQ-05)
|
|
|
|
| Application Field |
- Photoisomerization
- Photocatalytic reaction
- Biological chemistry
- Fading test
|
| etc... |
|
 |
F-number |
F/2.8 |
Reciprocal Linear Dispersion |
8.3nm/mm (550nm) |
Resolution |
1:2.6nm 2:4.3nm 3:8.6mn (Slit width switcing) |
Stray Light* |
546.1 +/-10nm< 3.0 x 10-4
546.1 +/-50nm< 4.5 x 10-5
|
Slit Width |
1: 0.3 x 3.0mm
2: 0.5 x 3.0mm
3: 1.0 x 3.0mm
|
*Condition :
Mercury Line 546.1nm, Grating 1200g/mm,
Blaze Wavelength 400nm
|
| Blaze Wavelength and Diffraction Efficiency |
|
|
|
Blaze Wavelength : 400nm (Recommended wavelength: 270 - 650nm) |
Blaze Wavelength : 250nm (Recommended wavelength: 220 - 450nm) |
|
|
Blaze Wavelength : 500nm (Recommended wavelength: 350 - 900nm) |
|
|
|
 |
Collimator lens reduces the divergence of light from the light guide and provide uniform light output. It is suitable for directional backlighting which requires clear silhouette of an object. |
 |
|
Description |
Mirror Module |
100W |
LAX-C100 |
UVB/UVA/VIS/IR |
300W |
MAX-303 |
UV/VIS/IR |
|
|